Continuous rating: max. 60 kV - 100 mA, 3 Kw
Control system: detection on secondary side, control on primary side.
Stability: within ±0.01% power supply variation ±10%
Security detection mechanism: (If any security item is detected,
X-ray turns OFF, indicating the error item on the CRT screen.)
Temperature of X-ray tube cooling water
Electric conductivity of X-ray tube cooling water
Flow rate of X-ray tube cooling water
Pressure of X-ray tube cooling water
Leak current of X-ray tube
Extreme high pressure, Extreme low pressure, Overload
Burning of X-ray indication lamp
Simultaneously opened X-ray shutter and lid
Tube voltage setting: 20 to 60 kV (at a step of 5 kV)
Tube current setting: 2 mA, 5 to 100 mA (at a step of 5 mA)
X-ray indicator lamp: on the system operation panel
Automatic aging function: the following items are operable with keys on the CRT operation screen.
X-ray tube (AFX-77L-Rh)
Target: Rh
Maximum rating: 3Kw
Super light elements including B, C and Wsix (tungstem silicide) can be analyzed at high precision
Unit is setup with the following channels: Fixed channel: Ba (barium) Fixed channel: Sr (strontium) Fixed channel: Ti (titanium) Scanning channel
Sample chamber: Sample irradiation method: at the underneath side Sample rotation: 6rpm (fixed rotation independent of the power supply frequency, 50/60 Hz)
X-ray atmosphere: vacuum/ the air (normally measured in a vacuum condition)
Maximum sample size: 200 mm
|